01 Overview

Rotating anode X-ray tube unit specifically designed for 100 mm diameter inserts
for heavy duty radiographic procedures.

02 Characteristics

IAE Housing High Voltage Heat Storage Capacity (kJ) Heat Storage Capacity (kHU) Max Cooling Rate Convection Max Cooling Rate Ventilation Net Weight Tech Data
C100 150 kV 1500 kJ 2000 kHU 250 W (20000 HU/min) 600 W (48000 HU/min) 25 kg C100
C100 G  – Special Version 150 kV 1500 kJ 2000 kHU 250 W (20000 HU/min) 600 W (48000 HU/min) 25 kg C100 G

• Lead lined aluminum body
• Two devices are present for thermal safety:
1 a bimetallic thermal switch, fitted externally on the anode end
2 a bimetallic thermal switch, internally assembled, series connected with stator common cable

 

Specifications
Overall length 520 mm
Maximum diameter 200 mm
Tube assembly net weight 25 kg
Nominal X-ray tube assembly voltage 150 kV
Maximum tube assembly heat content 1500 kJ  (2000 kHU)
Maximum continuous heat dissipation without fan 250 W (20000 HU/min)
Maximum continuous heat dissipation with fan 600 W (48000 HU/min)
Min. tube assembly inherent filtration 1.2 mm Al / 75 kV
Maximum leakage radiation at 1m from focal spots 0.44 mGy/h

 

INSERT FOR RELOADING
RTM101 H / HS
RTM102 H / HS
RTC600 HS
RTC602 HS
RTC700 HS
RTC1000 HS

03 Brochure