RTM102

01 Overview

Rotating anode X-ray tube specifically designed for general radiographic,
heavy duty X-ray systems.

02 Characteristics

IAE Tubes Focal Spots kW Rating Diam Angle Mat kJ kHU kVP Housing Tech Data TD CGR
RTM102 H 0.6/1.0 16/34 102 15° RTM 300 400 150 C52Super/C100/C100XT RTM102 H 0.6_1.0 RTM102 H 0.6_1.0 CGR
RTM102 HS 0.6/1.0 30/55 102 15° RTM 300 400 150 C52Super/C100/C100XT RTM102 HS 0.6_1.0 RTM102 HS 0.6_1.0 CGR
RTM102 H 0.6/1.2 16/42 102 15° RTM 300 400 150 C52Super/C100/C100XT RTM102 H 0.6_1.2 RTM102 H 0.6_1.2 CGR
RTM102 HS 0.6/1.2 29/75 102 15° RTM 300 400 150 C52Super/C100/C100XT RTM102 HS 0.6_1.2 RTM102 HS 0.6_1.2 CGR
RTM102 H 0.6/1.5 16/54 102 15° RTM 300 400 150 C52Super/C100/C100XT RTM102 H 0.6_1.5 RTM102 H 0.6_1.5 CGR
RTM102 HS 0.6/1.5 29/97 102 15° RTM 300 400 150 C52Super/C100/C100XT RTM102 HS 0.6_1.5 RTM102 HS 0.6_1.5 CGR
RTM102 H 1.0/2.0 34/64 102 15° RTM 300 400 150 C52Super/C100/C100XT RTM102 H 1.0_2.0 RTM102 H 1.0_2.0 CGR
RTM102 HS 1.0/2.0 55/125 102 15° RTM 300 400 150 C52Super/C100/C100XT RTM102 HS 1.0_2.0 RTM102 HS 1.0_2.0 CGR

 

• TZM target with large thermal capacity
• Bearing assembly and bearing lubricant system are
designed for long life under hardest x-ray
operating conditions
• Focal track in tungsten-rhenium alloy to prevent cracking
• High-thermal emittance oxide coating on anode
and rotor provides high anode heat dissipation rate
• Severe tests during conditioning assure best performances

 

Specifications
Maximum peak voltage 150 kV
Anode angle and diameter 15°- 102 mm
Maximum anode dissipation 1000 W (80000 HU/min)
Anode heat storage capacity 300 kJ (400 kHU)
Rotating anode speed 3000 – 10000 rpm

 

IAE HOUSINGS
C100
C52 Super
C100 XT

03 Brochure

RTM102