01Overview

Rotating anode X-ray tube for general radiographic and fluoro procedures,
from medium to heavy systems. Large anode angle allows greater X-ray coverage field.

02Characteristics

IAE TubesFocal SpotskW RatingDiamAngleMatkJkHUkVPHousingTech DataTD CGR
RTM92 H0.6/1.212/349015°RTM225300150C352/C52RTM92H 0.6_1.2RTM92 H 0.6_1.2 CGR
RTM92 HS0.6/1.222/659015°RTM225300150C352/C52RTM92 HS 0.6_1.2RTM92 HS 0.6_1.2 CGR
RTM92 H0.6/1.512/509015°RTM225300150C352/C52rRTM92 H 0.6_1.5RTM92 H 0.6_1.5 CGR
RTM92 HS0.6/1.522/909015°RTM225300150C352/C52RTM92 HS 0.6_1.5RTM92 HS 0.6_1.5 CGR
RTM92 H1.0/2.030/609015°RTM225300150C352/C52RTM92 H 1.0_2.0RTM92 H 1.0_2.0 CGR
RTM92 HS1.0/2.054/1109015°RTM225300150C352/C52rRTM92 HS 1.0_2.0RTM92 HS 1.0_2.0 CGR

 

• Bearing assembly and bearing lubricant system aredesigned for long life under hardest x-ray operating conditions
• Focal track in tungsten-rhenium alloy to prevent cracking
• High-thermal emittance oxide coating on anode and rotor provides high anode heat dissipation rate
• Severe tests during conditioning assure best performances
• Available also special versions for:
– reloading in GE-CGR and SIEMENS housings
(supplied with anode end adaptation piece)

 

Specifications
Maximum peak voltage150 kV
Anode angle and diameter15°- 90 mm
Maximum anode dissipation750 W (60000 HU/min)
Anode heat storage capacity225 kJ (300 kHU)
Rotating anode speed3000 – 10000 rpm

 

IAE Housings
C352
C52

 

Reloading
CG52 – CGR STATORIX 240
CG53 – CGR STATORIX 260
CP52 – PHILIPS ROTALIX 350
CC52 – COMET D09

03Brochure