01Overview

Rotating anode mammography X-ray tube with special bi-angled target,
for optimal performances with all techniques.
Replaces VARIAN M113SP and M113.

02Characteristics

IAE TubesFocal SpotskW RatingDiamAngleMatkJkHUkVPHousingTech Data
XM10160.1/0.30.7/4.98010°/16°Mo22530040C339C/C339V/C339E/C340VXM1016 0.1_1.3

 

• Two separeted focal tracks, small focus on 10° and large focus on 16°, optimal resolution performances
• Replacement for VARIAN M113SP and M113
• Minimum resolution 14×12 Ip/mm (width x lenght)
• Metal center section
• Molibdenum doped target
• Reduced thermal stress on the bearings improves tube life duration
• Severe tests during conditioning assure best performance

 

Specifications
Nominal X-ray tube voltage40 kV
Anode angle and diameterdouble angle target <br />10° /16° – 80
Maximum anode dissipation715 W (57200 HU/min)
Anode heat storage capacity225 kJ (300 kHU)
Rotating anode speed3000 – 10000 rpm

 

Max emission current
sF – 35 mA – 25 Kv
LF – 140 mA – 25 Kv

 

IAE HOUSINGS
C339E lateral HV and LV sockets, VARIAN version
C339C coaxial HV and LV sockets, COMET version
C339V coaxial HV and LV sockets, VARIAN version
C340V

03Brochure

XM1016

Download