IAE Tubes | Focal Spots | kW Rating | Diam | Angle | Mat | kJ | kHU | kVP | Housing | Tech Data |
---|---|---|---|---|---|---|---|---|---|---|
XM12T | 0.1/0.3 | 2/16 | 80 | 12.5° | RTM | 225 | 300 | 49 | C339C/C339V/C339E/C340V | XM12T 0.1_0.3 |
• Metal center section
• RTM target
• Reduced thermal stress on the bearings improves tube life duration
• Severe tests during conditioning assure best performance
Specifications | ||
---|---|---|
Nominal X-ray tube voltage | 49 kV | |
Anode angle and diameter | 12.5° – 80 mm | |
Maximum anode dissipation | 715 W (57200 HU/min) | |
Anode heat storage capacity | 225 kJ (300 kHU) | |
Rotating anode speed | 3000 – 10000 rpm |
Max emission current |
---|
sF – 35 mA – 25 kV |
LF – 140 mA – 25 kV |
IAE HOUSINGS |
---|
C339E lateral HV and LV sockets, VARIAN version |
C339C coaxial HV and LV sockets, COMET version |
C339V coaxial HV and LV sockets, VARIAN version |
C340V |