01Overview

Rotating anode X-ray tube unit specifically designed for 100 mm diameter inserts
for heavy duty radiographic procedures.

02Characteristics

IAE HousingHigh VoltageHeat Storage Capacity (kJ)Heat Storage Capacity (kHU)Max Cooling Rate ConvectionMax Cooling Rate VentilationNet WeightTech Data
C100150 kV1500 kJ2000 kHU250 W (20000 HU/min)600 W (48000 HU/min)25 kgC100
C100 G  – Special Version150 kV1500 kJ2000 kHU250 W (20000 HU/min)600 W (48000 HU/min)25 kgC100 G

• Lead lined aluminum body
• Two devices are present for thermal safety:
1 a bimetallic thermal switch, fitted externally on the anode end
2 a bimetallic thermal switch, internally assembled, series connected with stator common cable

 

Specifications
Overall length520 mm
Maximum diameter200 mm
Tube assembly net weight25 kg
Nominal X-ray tube assembly voltage150 kV
Maximum tube assembly heat content1500 kJ  (2000 kHU)
Maximum continuous heat dissipation without fan250 W (20000 HU/min)
Maximum continuous heat dissipation with fan600 W (48000 HU/min)
Min. tube assembly inherent filtration1.2 mm Al / 75 kV
Maximum leakage radiation at 1m from focal spots0.44 mGy/h

 

INSERT FOR RELOADING
RTM101 H / HS
RTM102 H / HS
RTC600 HS
RTC602 HS
RTC700 HS
RTC1000 HS

03Brochure