IAE Housing | High Voltage | Heat Storage Capacity (kJ) | Heat Storage Capacity (kHU) | Max Cooling Rate Convection | Net Weight | Tech Data |
---|---|---|---|---|---|---|
C100XS | 150 kV | 3500 kJ | 4660 kHU | 1400 W (112000 HU/min) with HE30+HE32 | 31 kg | C100 XS data |
• Lead lined aluminum body
• Three devices are present for thermal safety:
1 a bimetallic thermal switch, fitted externally on the anode end
2 a bimetallic thermal switch, internally assembled, series connected with stator common cable
3 a pressure microswitch, installed on cathode end, activated by bellow expansion
Specifications | ||
---|---|---|
Overall length | 515 mm | |
Maximum diameter | 200 mm | |
Tube assembly net weight | 31 kg | |
Nominal X-ray tube assembly voltage | 150 kV | |
Maximum tube assembly heat content | 3500 kJ (4660 kHU) | |
Maximum continuous heat dissipation | 1400 W (112000 HU/min) with HE30+HE32 | |
Min. tube assembly inherent filtration | 1.2 mm Al / 75 kV | |
Maximum leakage radiation at 1m from focal spots | 0.44 mGy/h |
INSERT FOR RELOADING |
---|
RTC600 HS |
RTC700 HS |
RTC1000 HS |