| IAE Housing | High Voltage | Heat Storage Capacity (kJ) | Heat Storage Capacity (kHU) | Max Cooling Rate Convection | Net Weight | Tech Data |
|---|---|---|---|---|---|---|
| C31 | 150 kV | 500 kJ | 670 kHU | 120 W (9600 HU/min) | 15 kg | C31 data |
• Lead lined aluminum body
• Three devices are present for thermal safety:
1 a bimetallic thermal switch, fitted externally on the anode end
2 a bimetallic thermal switch, internally assembled, series connected with stator common cable
3 a pressure switch
• H.T. cable socket: Parker type Compact Taper or Claymount type MINI75 4pin
• Solutions for higher dissipation available
| Specifications | ||
|---|---|---|
| Overall length | 400 mm | |
| Maximum diameter | 134 mm | |
| Tube assembly net weight | 15 kg | |
| Nominal X-ray tube assembly voltage | 150 kV | |
| Maximum tube assembly heat content | 500 kJ (670 kHU) | |
| Maximum continuous heat dissipation | 120 W (9600 HU/min) | |
| Maximum tube assembly inherent filtration | 1.2 mm Al / 75 kV | |
| Maximum leakage radiation at 1m from focal spots | 0.18 mGy/h |
| INSERT FOR RELOADING |
|---|
| X22 |
| RTM30 HS |
| RTM72 H/HS |
| RTM77 H |
| RTM75 H/HS |