https://www.iae.it/wp-content/uploads/2017/12/C32-Rev0.pdf
| IAE Housing | High Voltage | Heat Storage Capacity (kJ) | Heat Storage Capacity (kHU) | Max Cooling Rate Ventilation | Net Weight | Tech Data |
|---|---|---|---|---|---|---|
| C32 | 125 kV | 500 kJ | 670 kHU | with HE30-1000 W (80000 HU/min) with HE31-1200 W (96000 HU/min) | 15 kg | C32 data |
• Lead lined aluminum body
• Internal pump for oil circulation, to improve thermal exchange
• Three devices are present for thermal safety:
1 a bimetallic thermal switch, fitted externally on the anode end
2 a bimetallic thermal switch, internally assembled, series connected with stator common cable
3 a pressure switch
• H.T. cable socket: Parker type Compact Taper or Claymount type MINI75 4pin
• Solutions for higher dissipation avilable
| Specifications | ||
|---|---|---|
| Overall length | 435 mm | |
| Maximum diameter | 150 mm | |
| Tube assembly net weight | 15 kg | |
| Nominal X-ray tube assembly voltage | 125 kV | |
| Maximum tube assembly heat content | 500 kJ (670 kHU) | |
| Maximum continuous heat dissipation | 1000 W (80000 HU/min) with HE30 | |
| 1200 W (96000 HU/min) with HE31 | ||
| Min. tube assembly inherent filtration | 1.2 mm Al / 75 kV | |
| Maximum leakage radiation at 1m from focal spots | 0.22 mGy/h |
| INSERT FOR RELOADING |
|---|
| X20P |
| RTM70 H/HS |

