01Overview

Water cooled mammography tube unit
for beam scanning mammography equipments
and high patients throughput screening applications.

02Characteristics

IAE HousingHigh VoltageHeat Storage Capacity (kJ)Heat Storage Capacity (kHU)Max Cooling Rate ConvectionNet WeightTech Data
C340V49 kV375 kJ426 kHU800 W (64000 HU/min) with HE3016 kgC340 V data

• An external water circulation cooling jacket, in combination with an internal oil circulation system, ensures a heat dissipation three times higher than conventional mammography tube units.
• For thermal protection there is a bimetallic thermal switch internally assembled series connected with stator common cable

 

Specifications
Overall length340 mm
Maximum diameter450 mm
Tube assembly net weight16 kg
Nominal X-ray tube assembly voltage49 kV
Maximum tube assembly heat content375 kJ  (426 kHU)
Maximum continuous heat dissipation800 W (64000 HU/min) with HE30
Min. tube assembly inherent filtration0.5 mm-Be
Maximum leakage radiation at 1m from focal spots45 µmGy/h

 

INSERT FOR RELOADING
XM12
XM12T
XM15
XM15T
XM1016
XM1016T

03Brochure