| IAE Housing | High Voltage | Heat Storage Capacity (kJ) | Heat Storage Capacity (kHU) | Max Cooling Rate Convection | Net Weight | Tech Data |
|---|---|---|---|---|---|---|
| C340V | 49 kV | 375 kJ | 426 kHU | 800 W (64000 HU/min) with HE30 | 16 kg | C340 V data |
• An external water circulation cooling jacket, in combination with an internal oil circulation system, ensures a heat dissipation three times higher than conventional mammography tube units.
• For thermal protection there is a bimetallic thermal switch internally assembled series connected with stator common cable
| Specifications | ||
|---|---|---|
| Overall length | 340 mm | |
| Maximum diameter | 450 mm | |
| Tube assembly net weight | 16 kg | |
| Nominal X-ray tube assembly voltage | 49 kV | |
| Maximum tube assembly heat content | 375 kJ (426 kHU) | |
| Maximum continuous heat dissipation | 800 W (64000 HU/min) with HE30 | |
| Min. tube assembly inherent filtration | 0.5 mm-Be | |
| Maximum leakage radiation at 1m from focal spots | 45 µmGy/h |
| INSERT FOR RELOADING |
|---|
| XM12 |
| XM12T |
| XM15 |
| XM15T |
| XM1016 |
| XM1016T |