C340V

01 Overview

Water cooled mammography tube unit
for beam scanning mammography equipments
and high patients throughput screening applications.

02 Characteristics

IAE Housing High Voltage Heat Storage Capacity (kJ) Heat Storage Capacity (kHU) Max Cooling Rate Convection Net Weight Tech Data
C340V 49 kV 375 kJ 426 kHU 800 W (64000 HU/min) with HE30 16 kg C340 V data

• An external water circulation cooling jacket, in combination with an internal oil circulation system, ensures a heat dissipation three times higher than conventional mammography tube units.
• For thermal protection there is a bimetallic thermal switch internally assembled series connected with stator common cable

 

Specifications
Overall length 340 mm
Maximum diameter 450 mm
Tube assembly net weight 16 kg
Nominal X-ray tube assembly voltage 49 kV
Maximum tube assembly heat content 375 kJ  (426 kHU)
Maximum continuous heat dissipation 800 W (64000 HU/min) with HE30
Min. tube assembly inherent filtration 0.5 mm-Be
Maximum leakage radiation at 1m from focal spots 45 µmGy/h

 

INSERT FOR RELOADING
XM12
XM12T
XM15
XM15T
XM1016
XM1016T

03 Brochure

C340V

Download