01Overview

Rotating anode X-ray tube unit specifically designed for anode diameter inserts
ranging from 70 mm to 80 mm, for medium/high duty radiographic procedures.

02Characteristics

IAE HousingHigh VoltageHeat Storage Capacity (kJ)Heat Storage Capacity (kHU)Max Cooling Rate ConvectionMax Cooling Rate VentilationNet WeightTech Data
C40150 kV900 kJ1250 kHU180 W (14400 HU/min)16 kgC40 data

 

• Lead lined aluminum body
• Two devices are present for thermal safety:
1 a bimetallic thermal switch, internally assembled, series connected with stator common cable
2 a pressure microswitch, installed on cathode end, activated by bellow expansion

 

Specifications
Overall length490 mm
Maximum diameter152 mm
Tube assembly net weight16 kg
Nominal X-ray tube assembly voltage150 kV
Maximum tube assembly heat content900 kJ  (1250 kHU)
Maximum continuous heat dissipation180 W (14400 HU/min)
Min. tube assembly inherent filtration1.2 mm Al / 75 kV
Maximum leakage radiation at 1m from focal spots0.44 mGy/h

 

INSERT FOR RELOADING
X39
X42
X76

03Brochure