| IAE Housing | High Voltage | Heat Storage Capacity (kJ) | Heat Storage Capacity (kHU) | Max Cooling Rate Convection | Max Cooling Rate Ventilation | Net Weight | Tech Data |
|---|---|---|---|---|---|---|---|
| C52 Super | 150 kV | 1280 kJ | 1700 kHU | 230 W (18400 HU/min | 370 W (29600 HU/min) | 26 kg | C52 Super data |
• Lead lined aluminum body
• Three devices are present for thermal safety:
1 a bimetallic thermal switch, fitted externally on the anode end
2 a bimetallic thermal switch, internally assembled, series connected with stator common cable
3 a pressure microswitch, installed on cathode end, activated by bellow expansion
| Specifications | ||
|---|---|---|
| Overall length | 490 mm | |
| Maximum diameter | 182 mm | |
| Tube assembly net weight | 26 kg | |
| Nominal X-ray tube assembly voltage | 150 kV | |
| Maximum tube assembly heat content | 1280 kJ (1700 kHU) | |
| Maximum continuous heat dissipation without fan | 230 W (18400 HU/min) | |
| Maximum continuous heat dissipation with fan | 370 W (29600 HU/min) | |
| Min. tube assembly inherent filtration | 1.2 mm Al / 75 kV | |
| Maximum leakage radiation at 1m from focal spots | 0.44 mGy/h |
| INSERT FOR RELOADING |
|---|
| RTM90 H / HS |
| RTM92 H / HS |
| RTM101 H / HS |
| RTM102 H / HS |
| RTC600 HS |
| RTC700 HS |
| RTC1000 HS |