01 Overview

Rotating anode X-ray tube for general radiographic and fluoro procedures,
from medium to heavy systems. Large anode angle allows greater X-ray coverage field.

02 Characteristics

IAE Tubes Focal Spots kW Rating Diam Angle Mat kJ kHU kVP Housing Tech Data TD CGR
RTM92 H 0.6/1.2 12/34 90 15° RTM 225 300 150 C352/C52 RTM92H 0.6_1.2 RTM92 H 0.6_1.2 CGR
RTM92 HS 0.6/1.2 22/65 90 15° RTM 225 300 150 C352/C52 RTM92 HS 0.6_1.2 RTM92 HS 0.6_1.2 CGR
RTM92 H 0.6/1.5 12/50 90 15° RTM 225 300 150 C352/C52r RTM92 H 0.6_1.5 RTM92 H 0.6_1.5 CGR
RTM92 HS 0.6/1.5 22/90 90 15° RTM 225 300 150 C352/C52 RTM92 HS 0.6_1.5 RTM92 HS 0.6_1.5 CGR
RTM92 H 1.0/2.0 30/60 90 15° RTM 225 300 150 C352/C52 RTM92 H 1.0_2.0 RTM92 H 1.0_2.0 CGR
RTM92 HS 1.0/2.0 54/110 90 15° RTM 225 300 150 C352/C52r RTM92 HS 1.0_2.0 RTM92 HS 1.0_2.0 CGR

 

• Bearing assembly and bearing lubricant system aredesigned for long life under hardest x-ray operating conditions
• Focal track in tungsten-rhenium alloy to prevent cracking
• High-thermal emittance oxide coating on anode and rotor provides high anode heat dissipation rate
• Severe tests during conditioning assure best performances
• Available also special versions for:
– reloading in GE-CGR and SIEMENS housings
(supplied with anode end adaptation piece)

 

Specifications
Maximum peak voltage 150 kV
Anode angle and diameter 15°- 90 mm
Maximum anode dissipation 750 W (60000 HU/min)
Anode heat storage capacity 225 kJ (300 kHU)
Rotating anode speed 3000 – 10000 rpm

 

IAE Housings
C352
C52

 

Reloading
CG52 – CGR STATORIX 240
CG53 – CGR STATORIX 260
CP52 – PHILIPS ROTALIX 350
CC52 – COMET D09

03 Brochure

RTM92

Download