01 Overview

Rotating anode X-ray tube unit specifically designed for mobile C-arms equipment.
Suitable for max 72 mm anode diameter inserts.

02 Characteristics

IAE Housing High Voltage Heat Storage Capacity (kJ) Heat Storage Capacity (kHU) Max Cooling Rate Convection Max Cooling Rate Ventilation Net Weight Tech Data
C30 125 kV 500 kJ 670 kHU 140 W (11200 HU/min) with HE30-500 W (40000 HU/min)  15 kg C30 data

 

• Lead lined aluminum body
• Internal pump for oil circulation, to improve thermal exchange
• Three devices are present for thermal safety:
1 a bimetallic thermal switch, fitted externally on the anode end
2 a bimetallic thermal switch, internally assembled, series connected with stator common cable
3 a pressure switch
• H.T. cable socket: Parker type Compact Taper or Claymount type MINI75 4pin
• Solutions for higher dissipation available

 

Specifications
Overall length 416 mm
Maximum diameter 134 mm
Tube assembly net weight 15 kg
Nominal X-ray tube assembly voltage 125 kV
Maximum tube assembly heat content 500 kJ  (670 kHU)
Maximum continuous heat dissipation 140 W (11200 HU/min)
500 W (40000 HU/min) with HE30
Minimum tube assembly inherent filtration 1.2 mm Al / 75 kv
Maximum leakage radiation at 1m from focal spots  0.18 mGy/h

 

INSERT FOR RELOADING
X20 P
RTM70 H/HS
RTM75 H/HS

03 Brochure