01 Overview

Rotating anode X-ray tube unit specifically designed for anode diameter inserts ranging from 70 mm to 80 mm, for medium/duty radiographic procedures.

02 Characteristics

IAE Housing High Voltage Heat Storage Capacity (kJ) Heat Storage Capacity (kHU) Max Cooling Rate Convection Net Weight Tech Data
C33 125 kV 500 kJ 670 kHU 1000/1200 W 16 kg C33 data

 

• Lead lined aluminum body
• Two devices are present for thermal safety:
1 a bimetallic thermal switch, internal assembled, series connected with stator common cable
2 a pressure microswitch, installed on cathode end, activated by bellow expansion

 

Specifications
Overall length 435 mm
Maximum diameter 150 mm
Tube assembly net weight 16 kg
Nominal X-ray tube assembly voltage 125 kV
Maximum tube assembly heat content 500 kJ  (670 kHU)
Maximum continuous heat dissipation 1000 W (80000 HU/min) with HE30
1200 W (96000 HU/min) with HE31
Min. tube assembly inherent filtration 1.2 mm Al / 75 kV
Maximum leakage radiation at 1m from focal spots 0.22 mGy/h

 

INSERT FOR RELOADING
RTM780
RTM80

03 Brochure