01Overview

Rotating anode X-ray tube unit specifically designed for anode diameter inserts ranging from 70 mm to 80 mm, for medium/duty radiographic procedures.

02Characteristics

IAE HousingHigh VoltageHeat Storage Capacity (kJ)Heat Storage Capacity (kHU)Max Cooling Rate ConvectionNet WeightTech Data
C33125 kV500 kJ670 kHU1000/1200 W16 kgC33 data

 

• Lead lined aluminum body
• Two devices are present for thermal safety:
1 a bimetallic thermal switch, internal assembled, series connected with stator common cable
2 a pressure microswitch, installed on cathode end, activated by bellow expansion

 

Specifications
Overall length435 mm
Maximum diameter150 mm
Tube assembly net weight16 kg
Nominal X-ray tube assembly voltage125 kV
Maximum tube assembly heat content500 kJ  (670 kHU)
Maximum continuous heat dissipation1000 W (80000 HU/min) with HE30
1200 W (96000 HU/min) with HE31
Min. tube assembly inherent filtration1.2 mm Al / 75 kV
Maximum leakage radiation at 1m from focal spots0.22 mGy/h

 

INSERT FOR RELOADING
RTM780
RTM80

03Brochure